Measurement principles
How the light meets your sample
Every Hellma probe sends light into the process and returns it to the spectrometer. What differs is the path the light takes.
Espada probes
Fluorescence
How it works
- Excitation light travels down one fiber. A mirror turns it 90° so it crosses the measuring gap.
- Fluorescent molecules in the sample absorb it and emit light at longer wavelengths, in all directions.
- A mirror at the far side of the gap sends the excitation light back for a second pass, which generates more fluorescence.
- The collection fiber sits at 90° to the excitation beam. Emission heading away from it hits a mirror at the bottom of the gap and is reflected back up to the fiber.
Advantages
- The right-angle geometry keeps excitation light out of the collection path, so the background stays low.
- Two mirrors double the excitation pass and recover downward emission, for more signal.
- Quartz glass optics for the UV/VIS range.