Precious-metals refining: from dilution to ATR

Transition metals absorb UV-Vis light very strongly. For one refiner, that made a standard measurement almost unworkable until the probe changed.

The problem

A precious-metals refining company measured with a Falcata transflection probe with a 1 mm (0.04 in) light path. The samples absorbed so strongly that they had to be diluted by a factor of one million to bring the absorbance into a usable range.

The fix

They moved to an ATR probe. ATR has an extremely short path length, so strongly absorbing liquids stay in range, and it worked beautifully.

Why metals are hard

Transition metals such as copper, nickel, cobalt, iron, manganese and zinc absorb light very strongly, with bands spread from the UV into the near infrared.

What is measured
Transition-metal concentration
AbsorbanceusableTransflection1 mm (0.04 in) light pathfar off scaledilute 1 : 1,000,000ATRtiny penetration depthin range