Filter dryer end point
Filter dryers wash and dry pharmaceutical product before blending and tableting. Drying for a fixed time, or stopping to pull a sample, gives no continuous view of the moisture.
Measured in the bowl
Diffuse reflectance probes sit in measurement ports in the dryer bowl, placed to suit its size, and connect to an FT-NIR spectrometer by optical fibers. The NIR light penetrates 2–3 mm (0.08–0.12 in) into the product.
Keeping the tip clean
Product can build up on the probe tip, and the probe would then keep measuring the same material. Probes with a software-controlled purge blow the tip clear at regular intervals.
Calibration
Spectra taken during drying are matched to samples analyzed by loss on drying or Karl Fischer titration. A PLS model then reports moisture or solvent content in real time.
Based on Bruker Optics application note AF 505, Filter Drier Endpoint Analysis using FT-NIR Spectroscopy.